6100-Gfast Multifunction Copper Tester Operation Manual
© 2020 Radiodetection Ltd
59
2. Press the left/right arrow keys to view and
select the options.
OR
3. Press
on a value to open a list box of
options.
4. Press
to confirm the value.
Snapshot
The voltage
Snapshot
tab shows measured volts for
each lead combination.
The parameters are as follows:
Current Group
is the current test group name,
for example
Multimeter
.
Current Test
displays the test being
measured. Here you can switch to another test
from the given list. However, this action will
terminate the ongoing measurement test.
Selected Pair
allows you to select
All Pairs
or
a single wire pair for measurement.
Termination Resistance
defines how much
resistance the tester uses to connect to the line
and measure voltages. The default value of
100 kΩ is widely used in the industry.
Threshold Settings
button opens a new page
that allows you to set the AC/DC voltage
thresholds for each pair.
Result
displays a snapshot of the pass/fail
status, volts and frequency (Hz) values for
each continuous AC/DC pair.
Test Status
displays the current state of
measurement or an error message.
To select the parameter values:
1. Press the up/down arrow keys to highlight the
desired parameter.
2. Press the left/right arrow keys to view and
select the options.
OR
3. Press
on a value to open a list box of
options.
4. Press
to confirm the value.
9.3 Current
The
Current
test allows you to measure the AC/DC
currents and frequency in the loop. DSL service does
not directly use this measurement.
Snapshot
The current
Snapshot
shows measured AC/DC results
for each lead combination. DC current is traditionally
measured for POTS/voice with a minimum of 23 mA,
but less DC current can indicate line faults or too much
loss. The Current test defaults to the
Continuous DC
page from the
Multimeter
menu.
The parameters are as follows:
Current Group
is the current test group name,
for example
Multimeter
.
Current Test
displays the test being
measured. Here you can switch to another test
from the given list. However, this action will
terminate the ongoing measurement test.
Selected Pair
allows you to select
All Pairs
or
a single wire pair for measurement.