6100-Gfast Multifunction Copper Tester Operation Manual
© 2020 Radiodetection Ltd
67
Type
is the threshold limit for each pair: either
Maximum
or
Minimum
.
Enabled
is a flag to use the voltage threshold:
either
Yes
or
No
.
To set threshold values:
1. Press the up/down arrow keys to highlight the
desired parameter.
2. Press the left/right arrow keys to view and
select the options.
OR
3. Press
on a value to open a list box of
options or the alphanumeric editor screen and
use the navigation keys to scroll through.
4. Press
to confirm the value.
Continuous
The
Continuous
function measures each pair,
displaying and updating results in
real time
, until you
stop the test. The last valid results then remain
displayed.
Remember that the continuous application of a higher
voltage tends to dry up moisture in faults, making it
seem as if they have disappeared; but only the
moisture is gone and will likely reappear another time.
Note: The test should be run with an open circuit at the
far end.
Note: If
UFED Control
option is enabled, due to its
limitation, the
Continuous
function will act the
same as the test in
Snapshot
mode and
terminate after the
Soak Time
period expires.
Current Group
is the current test group name,
for example
Multimeter
.
Current Test
displays the test being
measured. Here you can switch to another test
from the given list. However, this action will
terminate the ongoing measurement test.
Selected Pair
allows you to select a single
wire pair for measurement.
Soak Voltage
allows you to apply a voltage
value to the pair(s) currently under test. If your
unit has the HIVOLT (high voltage) software
option configured, the
Soak Voltage
parameter
values are extended up to 500 V maximum.
Note: Since
Soak Voltage
is inherited in other tests
that use the
Isolation
test, the HIVOLT software
option also applies to the following tests:
Pair
Detective
and
User Auto Test
.
Minimal Soak Time
sets a soak period of time
from 1 to 60 seconds. When the test duration
reaches the
Minimal Soak Time
, the pass/fail
status is displayed and then periodically
refreshed, until you stop the test.
Threshold Settings
button opens a new page
that allows you to set isolation thresholds for
each pair.