6100-Gfast Multifunction Network Analyzer Operation Manual
72
© 2020 Radiodetection Ltd
Test Status
displays the current state of
measurement or an error message.
To select the Station Ground parameter values:
1. Press the up/down arrow keys to highlight the
desired parameter.
2. Press the left/right arrow keys to view and
select the options.
OR
3. Press
on a value to open a list box of
options.
4. Press
to confirm the value.
Threshold Settings
The
Test Thresholds
page allows you to set the
threshold values for the Ground Resistance. The
threshold is a programmable value you can use as an
on-screen indicator in each test when a measured
value is too high or too low.
Type
is the threshold limit: either
Maximum
or
Minimum
.
Enabled
is a flag to use the current threshold:
either
Yes
or
No
.
To set threshold values:
1. Press the up/down arrow keys to highlight the
desired parameter.
2. Press the left/right arrow keys to view and
select the options.
OR
3. Press
on a value to open a list box of
options or the alphanumeric editor screen and
use the navigation keys to scroll through.
4. Press
to confirm the value.
Result Details
The
Result Details
read-only page displays the
Station Ground
results and the pass/fail status.
Test Group
is the overall test group of the
selected saved results file, for example
Multimeter
.
Test Identification
is the test type of the
group, for example
Station Ground
.
Test Modes
identifies the manner in which the
test was run, for example
Snapshot
.
9.9 Stressed Balance
Stressed Balance
is an optional test that continuously
generates high voltage (135 VDC) to the cable pair
under test while measuring balance. The high voltage
stresses
the conductors in order to reveal imbalances