PROBLEM
POSSIBLE CAUSES
WHAT TO DO
•
Baud rate too high for cable
and terminal being used.
•
Change to a lower baud rate.
Make sure that the new 1256
baud rate matches the new
terminal baud rate (to set the
1256 baud rate, see
Chapter
4, GPIB and RS-232
Interfaces
, under the heading
Setting the Baud Rate
).
•
Cable is excessively long,
or not of high quality.
•
Use a high-quality, shielded
cable.
Text from 1256 to remote
terminal overwrites itself,
staying on one line.
•
Termination characters of
1256 not compatible with
terminal setup.
•
Set 1256 to use carriage return
and line feed (CR-LF) as
termination characters (see
Chapter 4, GPIB and RS-232
Interfaces
, under the heading
Setting the Message
Termination Characters
).
Text from 1256 to remote
terminal is double-spaced
(lines of text are separated by
blank lines).
•
Termination characters of
1256 not compatible with
terminal setup.
•
Set 1256 to use only carriage
return (CR) as termination
character (see
Chapter 4,
GPIB and RS-232 Interfaces
,
under the heading
Setting the
Message Termination
Characters
).
•
Echo mode of remote
terminal is not set to half-
duplex (local echo).
•
Set terminal for half-duplex
(local echo) mode. Consult the
terminal user’s manual for
instructions.
Remote terminal does not
display characters that are
typed.
•
HyperTerminal (included
with Microsoft Windows)
can fail to echo characters
properly.
•
Download newer version,
known as “HyperTerminal
Private Edition” from:
http://www.hilgraeve.com
Содержание 1256E
Страница 5: ......
Страница 6: ...This page was left intentionally blank...
Страница 20: ...1256E User Manual Introduction 1 6 Racal Instruments 2002 This page was left intentionally blank...
Страница 30: ...1256E User Manual Getting Started 2 10 Racal Instruments 2002 This page was left intentionally blank...
Страница 96: ...1256E User Manual Computer Interfaces 4 32 Racal Instruments 2002 This page was left intentionally blank...
Страница 104: ...1256E User Manual SCPI Command Basics 5 8 Racal Instruments 2002 Figure 5 1 IEEE 488 2 Status Reporting Model...
Страница 146: ...1256E User Manual SCPI Command Reference 6 6 Racal Instruments 2002...
Страница 179: ...1256E User Manual Racal Instruments 2002 Advanced SCPI Operations 8 15 select input and output trigger sources and so on...
Страница 190: ...1256E User Manual Product Support 9 4 Racal Instruments 2002 This page was left intentionally blank...