1256E User Manual
Racal Instruments
©
2002
Advanced SCPI Operations 8-9
Selecting the
Trigger Count
The trigger count determines how many elements will be scanned
in the scan list. That is, if the trigger count is 10, then the scan list
will advance 10 times before halting. Any triggers received from
the selected trigger source will be ignored after the trigger count
has been satisfied.
The syntax for the TRIGGER:COUNT command is:
:TRIGger
[:SEQuence]
:COUNt
<trigger
count>
The <trigger count> is a numeric value between 1 and
21747483647 (2
31
-1).
The present trigger count may be queried by using the
TRIGGER:COUNT? query.
Selecting a Trigger
Delay
The 1256 may be programmed to delay before acting on a trigger.
This provides the ability to slow down the scan list operation if so
desired. The TRIGGER:DELAY command is used to program the
trigger delay. The syntax for this command is:
:TRIGger
[:SEQuence]
:DELay
<trigger
delay>
The <trigger delay> is a real number between 0.0 and 10.0. The
trigger delay is programmed in seconds, with a resolution of 1
microsecond. This is the
resolution
and not the
accuracy
. That
means that you will receive 0.00000 (5 decimal places) when you
read it. Any delay over 10 milliseconds rounds the delay to the
nearest 10 milliseconds.
When a non-zero trigger delay is programmed, the 1256 performs
an “idle wait” for the trigger delay period before acting on the
trigger.
The present trigger delay may be queried by using the
TRIGGER:DELAY? command.
Содержание 1256E
Страница 5: ......
Страница 6: ...This page was left intentionally blank...
Страница 20: ...1256E User Manual Introduction 1 6 Racal Instruments 2002 This page was left intentionally blank...
Страница 30: ...1256E User Manual Getting Started 2 10 Racal Instruments 2002 This page was left intentionally blank...
Страница 96: ...1256E User Manual Computer Interfaces 4 32 Racal Instruments 2002 This page was left intentionally blank...
Страница 104: ...1256E User Manual SCPI Command Basics 5 8 Racal Instruments 2002 Figure 5 1 IEEE 488 2 Status Reporting Model...
Страница 146: ...1256E User Manual SCPI Command Reference 6 6 Racal Instruments 2002...
Страница 179: ...1256E User Manual Racal Instruments 2002 Advanced SCPI Operations 8 15 select input and output trigger sources and so on...
Страница 190: ...1256E User Manual Product Support 9 4 Racal Instruments 2002 This page was left intentionally blank...