Switching and
Heat Dissipation
The 1256 accommodates all Adapt-a-Switch plug-ins, which gives
it exceptionally high-density switching capability. When switching a
large quantity of signals near its maximum rated loading current, a
plug-in generates heat due to the resistance of its conducting
paths and relay contacts. To prevent excessive heat, certain
guidelines apply regarding the maximum number of channels that
may carry full rated current at one time. In a high-density switching
system, it is not reasonable to have all relays simultaneously
closed, all carrying full rated current, operating at the maximum
rated ambient temperature. Some form of de-rating is necessary.
Whether or not a plug-in requires de-rating, and to what degree,
depends on the type of plug-in, its switching density, and the
ambient temperature. The following guidelines provide practical
de-rating for plug-ins used in the 1256 chassis.
Signal Switch Plug-Ins
Signal switch plug-ins consist of low- to moderate-current modules
such as the 1260-117 and 1260-118. For these plug-ins, up to
25% of the channels on each plug-in may be simultaneously
closed under full-rated current if the ambient temperature is 50
o
C
or less. If the ambient temperature is greater than 50
o
C, de-rate
the current by 20%.
Power Switch Plug-Ins
Power switch plug-ins are high-current modules such as the 1260-
120 and 1260-121. For these, their full rated current is allowed on
all channels simultaneously at ambient temperatures up to 50
o
C.
For ambient temperatures above 50
o
C, de-rate the current by 20%.
Содержание 1256E
Страница 5: ......
Страница 6: ...This page was left intentionally blank...
Страница 20: ...1256E User Manual Introduction 1 6 Racal Instruments 2002 This page was left intentionally blank...
Страница 30: ...1256E User Manual Getting Started 2 10 Racal Instruments 2002 This page was left intentionally blank...
Страница 96: ...1256E User Manual Computer Interfaces 4 32 Racal Instruments 2002 This page was left intentionally blank...
Страница 104: ...1256E User Manual SCPI Command Basics 5 8 Racal Instruments 2002 Figure 5 1 IEEE 488 2 Status Reporting Model...
Страница 146: ...1256E User Manual SCPI Command Reference 6 6 Racal Instruments 2002...
Страница 179: ...1256E User Manual Racal Instruments 2002 Advanced SCPI Operations 8 15 select input and output trigger sources and so on...
Страница 190: ...1256E User Manual Product Support 9 4 Racal Instruments 2002 This page was left intentionally blank...