1256E User Manual
Racal Instruments
©
2002
SCPI Command Basics 5-25
be specified using either the long form or the short form.
Whenever a discrete parameter is used, the query form of the
command returns the SHORT form of the parameter value, in
upper-case characters. That is, the command may be specified
using either “IMMEDIATE” or “IMM”, but the query:
TRIGGER:SOURCE?
will return the reply
IMM
NOT
IMMEDIATE
Boolean Parameters
Boolean parameters represent a single binary condition that is
either true or false. The 1256 accepts "OFF" or "0" for a false
condition. The 1256 accepts "ON" or "1" for a true condition. The
following command uses a boolean parameter:
[:ROUTe]
:MONitor
[:STATe] { OFF | ON | 0 | 1 }
The following commands turn the monitor OFF:
ROUTE:MONITOR:STATE OFF
ROUTE:MONITOR:STATE
0
MON
OFF
MON
0
The following commands turn the monitor ON
ROUTE:MONITOR:STATE ON
ROUT:MON:STAT
ON
MON
1
When Boolean parameters are queried, the 1256 always replies
with a “1", if the state is on, or “0", if the state is off. The keywords
“ON” and “OFF” are NOT returned in the reply to
ROUTE:MONITOR:STATE?, or any other query.
Command Input
The 1256 uses a command buffer to store commands sent either
Содержание 1256E
Страница 5: ......
Страница 6: ...This page was left intentionally blank...
Страница 20: ...1256E User Manual Introduction 1 6 Racal Instruments 2002 This page was left intentionally blank...
Страница 30: ...1256E User Manual Getting Started 2 10 Racal Instruments 2002 This page was left intentionally blank...
Страница 96: ...1256E User Manual Computer Interfaces 4 32 Racal Instruments 2002 This page was left intentionally blank...
Страница 104: ...1256E User Manual SCPI Command Basics 5 8 Racal Instruments 2002 Figure 5 1 IEEE 488 2 Status Reporting Model...
Страница 146: ...1256E User Manual SCPI Command Reference 6 6 Racal Instruments 2002...
Страница 179: ...1256E User Manual Racal Instruments 2002 Advanced SCPI Operations 8 15 select input and output trigger sources and so on...
Страница 190: ...1256E User Manual Product Support 9 4 Racal Instruments 2002 This page was left intentionally blank...