1256E User Manual
Computer Interfaces 4-22
Racal Instruments
©
2002
NOTE
Menu preferences are not normally retained after the
power is removed from the instrument. To make (a)
menu preference(s) permanent, make the desired
change(s) to settings first and then proceed to the
section Storing the Menu Preferences to make the
changes permanent. Following this procedure will put
the settings into non-volatile memory and allow them
to be restored during instrument power-up.
Setting the Message
Termination
Characters
Each message to or from the 1256 must end with an end-of-line
(EOL) character known as a terminator. You may select the EOL
terminator to be any one of the following:
•
Carriage return (CR) only
•
Line feed (LF) only (for non-terminal-based applications, such
as controlling via LabVIEW programs)
•
Carriage return and line feed (CR-LF)
The factory default is line feed only.
To select the EOL terminator, proceed as follows:
1. Press the top key. Rotate the knob until the top line reads:
>-----SETTINGS-----<
2. Select line 2 (next to top line) by pressing the second key. This
selects line 2 so that you may choose the type of setting to
work with:
-----SETTINGS-----
> Type: Display <
Option: Format
Status: Decimal
3. Rotate the knob until line 2 reads as shown below (lines 3 and
4 may differ):
-----SETTINGS-----
> Type: RS-232 <
Option: Data Bits
Status: 8
4. Press the third key to select line 3 of the display. Rotate the
knob until line 3 appears as shown below:
-----SETTINGS-----
Type: RS-232
> Option: EOL Mode <
Содержание 1256E
Страница 5: ......
Страница 6: ...This page was left intentionally blank...
Страница 20: ...1256E User Manual Introduction 1 6 Racal Instruments 2002 This page was left intentionally blank...
Страница 30: ...1256E User Manual Getting Started 2 10 Racal Instruments 2002 This page was left intentionally blank...
Страница 96: ...1256E User Manual Computer Interfaces 4 32 Racal Instruments 2002 This page was left intentionally blank...
Страница 104: ...1256E User Manual SCPI Command Basics 5 8 Racal Instruments 2002 Figure 5 1 IEEE 488 2 Status Reporting Model...
Страница 146: ...1256E User Manual SCPI Command Reference 6 6 Racal Instruments 2002...
Страница 179: ...1256E User Manual Racal Instruments 2002 Advanced SCPI Operations 8 15 select input and output trigger sources and so on...
Страница 190: ...1256E User Manual Product Support 9 4 Racal Instruments 2002 This page was left intentionally blank...