1256E User Manual
Computer Interfaces 4-20
Racal Instruments
©
2002
3. Rotate the knob until line 2 reads as shown below (lines 3 and
4 may differ):
-----SETTINGS-----
> Type: RS-232 <
Option: Baud Rate
Status: 9600
4. Press the third key to select line 3 of the display. Rotate the
knob until line 3 indicates the “Flow Control” option as shown
below:
-----SETTINGS-----
Type: Remote
> Option: FlowCtrl <
Status: Disabled
5. The three available flow control options on line 4 are:
•
Disabled
•
XON/XOFF
•
RTS/CTS
If line 4 of the display already indicates the desired choice,
then no action is necessary. Otherwise, press the fourth key.
Rotate the knob until the desired selection shows on line 4. In
this example, XON/XOFF is desired:
-----SETTINGS-----
Type: RS-232
Option: FlowCtrl
<Update?: XON/XOFF
Then press the fourth key to confirm the update. Pressing
any other button cancels the change.
NOTE
Menu preferences are not normally retained after the
power is removed from the instrument. To make (a)
menu preference(s) permanent, make the desired
change(s) to settings first and then proceed to the
section Storing the Menu Preferences to make the
changes permanent. Following this procedure will put
the settings into non-volatile memory and allow them
to be restored during instrument power-up.
Содержание 1256E
Страница 5: ......
Страница 6: ...This page was left intentionally blank...
Страница 20: ...1256E User Manual Introduction 1 6 Racal Instruments 2002 This page was left intentionally blank...
Страница 30: ...1256E User Manual Getting Started 2 10 Racal Instruments 2002 This page was left intentionally blank...
Страница 96: ...1256E User Manual Computer Interfaces 4 32 Racal Instruments 2002 This page was left intentionally blank...
Страница 104: ...1256E User Manual SCPI Command Basics 5 8 Racal Instruments 2002 Figure 5 1 IEEE 488 2 Status Reporting Model...
Страница 146: ...1256E User Manual SCPI Command Reference 6 6 Racal Instruments 2002...
Страница 179: ...1256E User Manual Racal Instruments 2002 Advanced SCPI Operations 8 15 select input and output trigger sources and so on...
Страница 190: ...1256E User Manual Product Support 9 4 Racal Instruments 2002 This page was left intentionally blank...