SCPI Command Basics 5-43
For example, the command:
SYST:KLOCK ON
will cause the 1256 to display a remote operation screen on the
display with a message indicating that local mode is in a lockout
condition.
The command:
SYST:KLOCK OFF
will also display a remote operation screen, however, the fourth
line will indicate a local option the user can activate by pressing
the fourth line soft key.
To query the present lockout state:
SYST:KLOCK?
will reply with either ON or OFF depending on the present state of
the 1256.
The 1256 operates in either automatic or manual user-interface
lockout mode. In either case, the 1256 will always power-up in
local mode. In manual lockout mode, the SYSTEM:KLOCK ON
command must be explicitly issued to prevent local operations. In
automatic lockout mode, any remote command detected by the
1256 puts the instrument in remote mode with lockout mode
automatically enabled. To allow local access, the 1256 must
receive a SYSTEM:KLOCK OFF command or experience a power
cycle event.
To make a lockout election from the user interface, select
-----SETTINGS-----
on the top most line, select
Type: Display
on the second line, and
Option: Lockout
from the third line and then choose the appropriate setting. The
default power-up state of the instrument is automatic mode
disabled. To elect a different power-up state, this setting like all
other menu options can be saved to non-volatile memory. For
further details on front-panel usage refer to
Chapter 3: USING
THE FRONT PANEL CONTROLS
.
Non-Volatile Update
The syntax for this command is shown below:
:SYSTem
:NVUPD
:NVUPD?
The SYSTEM:NVUPD and SYSTEM:NVUPD? query are used to
Содержание 1256E
Страница 5: ......
Страница 6: ...This page was left intentionally blank...
Страница 20: ...1256E User Manual Introduction 1 6 Racal Instruments 2002 This page was left intentionally blank...
Страница 30: ...1256E User Manual Getting Started 2 10 Racal Instruments 2002 This page was left intentionally blank...
Страница 96: ...1256E User Manual Computer Interfaces 4 32 Racal Instruments 2002 This page was left intentionally blank...
Страница 104: ...1256E User Manual SCPI Command Basics 5 8 Racal Instruments 2002 Figure 5 1 IEEE 488 2 Status Reporting Model...
Страница 146: ...1256E User Manual SCPI Command Reference 6 6 Racal Instruments 2002...
Страница 179: ...1256E User Manual Racal Instruments 2002 Advanced SCPI Operations 8 15 select input and output trigger sources and so on...
Страница 190: ...1256E User Manual Product Support 9 4 Racal Instruments 2002 This page was left intentionally blank...