SCPI Command Basics 5-19
*TST? Query
The *TST? query is supported only to be compliant with the IEEE-
488.2 standard. The 1256 performs an extensive self-test at every
power-up. If the system detects any failures, the condition is
indicated directly on the front-panel display. Normally self-test
errors are considered to be fatal conditions and result in a system
halt. In this condition, communication activity ceases. For this
reason, communications only function if self-test passes and
therefore a *TST? will always report a value of “0” indicating a
successful self-test.
*RCL Command
The *RCL command will recall the relay states from non-volatile
memory. The *RCL command may specify a non-volatile memory
location from which to recall the instrument state. That is, both of
the following formats are accepted:
*RCL
*RCL
<location>
If <location> is specified, it must be in the range 0 to 100. If
<location> is not specified, it will default to 100.
Location “0” has a special purpose. This location is used by the
1256 at power-up to set the initial relay states.
The *RCL command recalls the states of all relays in the system.
The relay states are stored using the *SAV command.
Note that path names, module names, status registers, include
lists, and so on are NOT affected by the *RCL command.
*SAV Command
The *SAV command will prepare to store a given state into non-
volatile memory. The actual process of updating non-volatile
memory for this command requires a separate SYSTEM:NVUPD
command. This allows multiple relay states, path names, module
names, relay verification flags, and verification masks to be
updated in rapid succession, followed by a single, slower update of
the actual non-volatile memory.
The *SAV command accepts one of two formats:
*SAV
*SAV
<location>
Содержание 1256E
Страница 5: ......
Страница 6: ...This page was left intentionally blank...
Страница 20: ...1256E User Manual Introduction 1 6 Racal Instruments 2002 This page was left intentionally blank...
Страница 30: ...1256E User Manual Getting Started 2 10 Racal Instruments 2002 This page was left intentionally blank...
Страница 96: ...1256E User Manual Computer Interfaces 4 32 Racal Instruments 2002 This page was left intentionally blank...
Страница 104: ...1256E User Manual SCPI Command Basics 5 8 Racal Instruments 2002 Figure 5 1 IEEE 488 2 Status Reporting Model...
Страница 146: ...1256E User Manual SCPI Command Reference 6 6 Racal Instruments 2002...
Страница 179: ...1256E User Manual Racal Instruments 2002 Advanced SCPI Operations 8 15 select input and output trigger sources and so on...
Страница 190: ...1256E User Manual Product Support 9 4 Racal Instruments 2002 This page was left intentionally blank...