If <location> is specified, it must be in the range from 0 through
100. If <location> is not specified, the instrument state will be
saved into non-volatile memory location 100.
State 0 is recalled at power-up.
The 1256 is shipped without any
data in state 0. This effectively tells the 1256 to open all relays
at power-up
. This default may be overwritten by placing all relays
in the desired power-up state, and then executing the command
*SAV 0
In the manner indicated earlier, a separate SYSTEM:NVUPD
command must follow to actually update the non-volatile memory.
If new relay modules are added to the system after the *SAV 0
command has been executed, the new relay modules will not be
programmed at power-up. Also, if module addresses are changed
after the execution of the *SAV 0 command, the modules whose
addresses have changed will not be programmed.
In general, whenever new modules are added, or module
addresses are changed, then the *SAV 0 command should be
used to place the relays of the new modules into the desired
power-up state.
*SPE Command
The status byte is returned to the controller in response to a Serial
Poll Enable (SPE) message.
The status byte carries the following information:
Bit Number
Bit Value
Description
1 (LSB)
1
Unused
2
2
Unused
3
4
Unused
4
8
Scan list at break point
5
16
Ready
6
32
Programming error
7
64
Service requested
8 (MSB)
128
Unused
Содержание 1256E
Страница 5: ......
Страница 6: ...This page was left intentionally blank...
Страница 20: ...1256E User Manual Introduction 1 6 Racal Instruments 2002 This page was left intentionally blank...
Страница 30: ...1256E User Manual Getting Started 2 10 Racal Instruments 2002 This page was left intentionally blank...
Страница 96: ...1256E User Manual Computer Interfaces 4 32 Racal Instruments 2002 This page was left intentionally blank...
Страница 104: ...1256E User Manual SCPI Command Basics 5 8 Racal Instruments 2002 Figure 5 1 IEEE 488 2 Status Reporting Model...
Страница 146: ...1256E User Manual SCPI Command Reference 6 6 Racal Instruments 2002...
Страница 179: ...1256E User Manual Racal Instruments 2002 Advanced SCPI Operations 8 15 select input and output trigger sources and so on...
Страница 190: ...1256E User Manual Product Support 9 4 Racal Instruments 2002 This page was left intentionally blank...