1256E User Manual
SCPI Command Basics 5-14
Racal Instruments
©
2002
•
The
Questionable Status Event Register
is cleared to 0. It is
read using the STATUS:QUESTIONABLE:EVENT? Query.
*ESE Command
The *ESE command sets the value of the
Standard Event Status
Enable Register
. The value of this register is logically ANDed with
the contents of the
Standard Event Status Register
(see the *ESR?
query description). If any bits of this AND operation are set, then
bit 5 of the Status Byte is set. Bit 5 is known as the Event
Summary Bit, or ESB, and is shown in Figure 2-3.
ESB of Status Byte =
(bit 0 of ESE AND bit 0 of ESR)
OR
(bit 1 of ESE AND bit 1 of ESR)
OR
(bit 2 of ESE AND bit 2 of ESR)
OR
(bit 3 of ESE AND bit 3 of ESR)
OR
(bit 4 of ESE AND bit 4 of ESR)
OR
(bit 5 of ESE AND bit 5 of ESR)
OR
(bit 7 of ESE AND bit 7 of ESR)
where:
ESE is the value as set by the *ESE command
ESR is the value which may be read with *ESR? query
The *ESE command has the format:
*ESE <ESE value>
where the “<ESE value>” is an integer numeric value in the range
from 0 through 255.
Содержание 1256E
Страница 5: ......
Страница 6: ...This page was left intentionally blank...
Страница 20: ...1256E User Manual Introduction 1 6 Racal Instruments 2002 This page was left intentionally blank...
Страница 30: ...1256E User Manual Getting Started 2 10 Racal Instruments 2002 This page was left intentionally blank...
Страница 96: ...1256E User Manual Computer Interfaces 4 32 Racal Instruments 2002 This page was left intentionally blank...
Страница 104: ...1256E User Manual SCPI Command Basics 5 8 Racal Instruments 2002 Figure 5 1 IEEE 488 2 Status Reporting Model...
Страница 146: ...1256E User Manual SCPI Command Reference 6 6 Racal Instruments 2002...
Страница 179: ...1256E User Manual Racal Instruments 2002 Advanced SCPI Operations 8 15 select input and output trigger sources and so on...
Страница 190: ...1256E User Manual Product Support 9 4 Racal Instruments 2002 This page was left intentionally blank...