1256E User Manual
Racal Instruments
©
2002
SCPI Command Basics 5-17
*IDN? Query
This query requests the instrument to identify itself. The EMS
responds to this query with the following reply:
Racal Instruments 1256 Switch System,<revision>
This reply indicates the manufacturer of the instrument is Racal
Instruments, that it is a 1256 Switch System, and the current
firmware revision. The firmware revision is a numeric, floating point
value. An example firmware revision is “3.10". A sample is shown
below:
Racal Instruments 1256 Switch System,3.10
*RST Command
The *RST command resets the instrument to its power-on default
state. These settings are shown in
Table 5-1
of this manual.
This command does
NOT
change the value of SCPI Operation or
Questionable status registers or IEEE-488.2 status registers,
condition registers, or enable registers. This command does
NOT
clear the error message queue, the input command buffer, or the
output reply buffer.
Содержание 1256E
Страница 5: ......
Страница 6: ...This page was left intentionally blank...
Страница 20: ...1256E User Manual Introduction 1 6 Racal Instruments 2002 This page was left intentionally blank...
Страница 30: ...1256E User Manual Getting Started 2 10 Racal Instruments 2002 This page was left intentionally blank...
Страница 96: ...1256E User Manual Computer Interfaces 4 32 Racal Instruments 2002 This page was left intentionally blank...
Страница 104: ...1256E User Manual SCPI Command Basics 5 8 Racal Instruments 2002 Figure 5 1 IEEE 488 2 Status Reporting Model...
Страница 146: ...1256E User Manual SCPI Command Reference 6 6 Racal Instruments 2002...
Страница 179: ...1256E User Manual Racal Instruments 2002 Advanced SCPI Operations 8 15 select input and output trigger sources and so on...
Страница 190: ...1256E User Manual Product Support 9 4 Racal Instruments 2002 This page was left intentionally blank...