196 .
LS 55 User’s Guide
NOTE:
The fan can be operated from either the 100V or 120V terminal. If an
operating voltage of 100V is selected the red fan wire should be connected
to the 120V terminal and vice versa.
NOTE:
The neon lamp can be operated from either the 220V or 240V terminal. If
an operating voltage of 220V is selected the orange lamp wire should be
connected to the 240V terminals and vice versa.
Changing the mains fuse
Having changed the operating voltage of the instrument the mains fuse must be
changed to correspond with the operating voltage. The table below gives the
correct fuse rating for the operating voltage.
Operating Voltage
Mains Fuse Rating
PerkinElmer Part Number
220-240 V 50/60 Hz
1A Slo Blo
OC97 3107
100-120 V 50/60 Hz
2A Slo Blo
OC97 3134
The mains fuse is located on the mains input panel at the rear of the instrument, as
shown in Figure 139.
Figure 139 The position of the mains fuse
1.
Undo the screw in the end of the fuse holder.
2.
Pull the fuse out.
3.
Fit the replacement fuse.
4.
Refit the screw.
Fuse holder
Содержание LS 55
Страница 1: ...LS 55 User s Guide ...
Страница 7: ...Warnings and Safety Information ...
Страница 12: ...12 LS 55 User s Guide ...
Страница 13: ...Introduction ...
Страница 19: ...Introduction 19 Supply Voltages 100 V 120 V 220 V 230 V Frequency 50 60 Hz Power 150 VA ...
Страница 20: ...20 LS 55 User s Guide ...
Страница 21: ...Unpacking and Installation ...
Страница 36: ...36 LS 55 User s Guide ...
Страница 37: ...Accessories Installation and Use ...
Страница 72: ...72 LS 55 User s Guide Figure 54 Using the powder accessory ...
Страница 93: ...Accessories Installation and Use 93 Figure 86 Masking the outer areas ...
Страница 121: ...RS232 Commands ...
Страница 163: ...System Description and Maintenance ...
Страница 165: ...System Description and Maintenance 165 Figure 112 The optical layout ...
Страница 198: ...198 LS 55 User s Guide ...
Страница 199: ...Index ...