System Description and Maintenance . 169
Scan Speed
nm/min
Block Average Width
(50 Hz)
Block Average Width
(60 Hz)
30 50
60
120 12
15
480 3
3
1500 1
1
Time Drive
In the time drive mode the amount of block averaging is determined by the value
of the response time specified. The amount of block averaging is adjusted (taking
into consideration the operating frequency of the instrument) to produce a
2% to
98% response to step signals over the specified
response time. The default
response time is 0.5 seconds.
Ratioing and Correction
After the sample and reference channels have been block averaged, they are
ratioed to improve the signal to noise and to remove source related artifacts. This
ratioed output is then scaled to either the wavelength range scanned or the time
scale chosen to give the corrected result. The results are multiplied by the
rhodamine correction curve, which is stored in the instrument, to correct for the
spectral response of the photomultiplier tubes and the transmission response of the
beamsplitter.
Содержание LS 55
Страница 1: ...LS 55 User s Guide ...
Страница 7: ...Warnings and Safety Information ...
Страница 12: ...12 LS 55 User s Guide ...
Страница 13: ...Introduction ...
Страница 19: ...Introduction 19 Supply Voltages 100 V 120 V 220 V 230 V Frequency 50 60 Hz Power 150 VA ...
Страница 20: ...20 LS 55 User s Guide ...
Страница 21: ...Unpacking and Installation ...
Страница 36: ...36 LS 55 User s Guide ...
Страница 37: ...Accessories Installation and Use ...
Страница 72: ...72 LS 55 User s Guide Figure 54 Using the powder accessory ...
Страница 93: ...Accessories Installation and Use 93 Figure 86 Masking the outer areas ...
Страница 121: ...RS232 Commands ...
Страница 163: ...System Description and Maintenance ...
Страница 165: ...System Description and Maintenance 165 Figure 112 The optical layout ...
Страница 198: ...198 LS 55 User s Guide ...
Страница 199: ...Index ...