Accessories: Installation and Use . 115
5.
Wipe the outside of the tube to dry it, and then insert the tube gently down
through the opening in the sample holder block.
6.
In Fl WinLab, start the TimeDrive application and enter the following
parameters:
•
Duration: 10 minutes
•
Data interval: 0.1 second
•
Excitation wavelength 302nm
•
Emission wavelength 445nm
•
Slits 5nm/5nm Ex/Em
NOTE:
These parameters are only for using coronene as the sample. If another
sample is used, then the excitation and emission wavelengths will have to
be set accordingly.
7.
Start the TimeDrive by clicking the Green traffic light.
8.
Locate the tube center in the focus crossover point to find the highest
intensity.
It is useful to look down into the sample tube. The fluorescence emission is
in the visible region, allowing you to to see when the sample is being excited
optimally.
NOTE:
There are two positions that produce high signals. The first is that which
correctly locates the center of the sample tube in the crossover point of the
excitation and emission light beams. This is the desired position. The
second position is that which causes light to be reflected directly from the
outer surface of the sample tube. This signal is unfortunately often higher
than the fluorescence signal. Determine that the position achieved after
alignment is correct by exchanging the sample tube for one containing only
solvent. If the signal changes dramatically, then the position is optimized
for fluorescence. If the signal does not change significantly, then the
position is optimized for scattered light, requiring further alignment.
Содержание LS 55
Страница 1: ...LS 55 User s Guide ...
Страница 7: ...Warnings and Safety Information ...
Страница 12: ...12 LS 55 User s Guide ...
Страница 13: ...Introduction ...
Страница 19: ...Introduction 19 Supply Voltages 100 V 120 V 220 V 230 V Frequency 50 60 Hz Power 150 VA ...
Страница 20: ...20 LS 55 User s Guide ...
Страница 21: ...Unpacking and Installation ...
Страница 36: ...36 LS 55 User s Guide ...
Страница 37: ...Accessories Installation and Use ...
Страница 72: ...72 LS 55 User s Guide Figure 54 Using the powder accessory ...
Страница 93: ...Accessories Installation and Use 93 Figure 86 Masking the outer areas ...
Страница 121: ...RS232 Commands ...
Страница 163: ...System Description and Maintenance ...
Страница 165: ...System Description and Maintenance 165 Figure 112 The optical layout ...
Страница 198: ...198 LS 55 User s Guide ...
Страница 199: ...Index ...