System Description and Maintenance . 195
Figure 138 Connections to the transformer
The wire connections to the terminals are as follows:
Wire Colour
Terminal
Line
Brown
100V, 120V, 220V or 240V
Fan Blue
0V
Fan
Red
120V or 100V
Neon Lamp
Orange
220V or 240V
Neon Lamp
Black
0V
Change the connections on the transformer as follows:
3.
Disconnect the terminal (brown) wire from the current terminal.
This will be connected to either the 240V, 220V, 120V or 100V terminal.
4.
Connect the brown wire to the terminal marked with the required operating
voltage.
NOTE:
The 240V
connection should be used for 230V supplies.
5.
Ensure that the operating voltage plate shows the correct operating voltage.
The appropriate operating voltage may be displayed on the reverse of the
current plate, or on the spare plate shipped with the instrument (L2251188
Mains Label 100 V / 220 V).
Содержание LS 55
Страница 1: ...LS 55 User s Guide ...
Страница 7: ...Warnings and Safety Information ...
Страница 12: ...12 LS 55 User s Guide ...
Страница 13: ...Introduction ...
Страница 19: ...Introduction 19 Supply Voltages 100 V 120 V 220 V 230 V Frequency 50 60 Hz Power 150 VA ...
Страница 20: ...20 LS 55 User s Guide ...
Страница 21: ...Unpacking and Installation ...
Страница 36: ...36 LS 55 User s Guide ...
Страница 37: ...Accessories Installation and Use ...
Страница 72: ...72 LS 55 User s Guide Figure 54 Using the powder accessory ...
Страница 93: ...Accessories Installation and Use 93 Figure 86 Masking the outer areas ...
Страница 121: ...RS232 Commands ...
Страница 163: ...System Description and Maintenance ...
Страница 165: ...System Description and Maintenance 165 Figure 112 The optical layout ...
Страница 198: ...198 LS 55 User s Guide ...
Страница 199: ...Index ...