142 .
LS 55 User’s Guide
Data Collection/Processing Set Up Commands
Command:
EXCITATION CORRECTION SELECTION
Syntax:
$CX code
Description:
Selects the excitation correction to be applied. Code is integer
with following values: 0 no correction, 1 correction. Power-on
default is 1.
NOTE:
Scan and time drive commands return error code 0111 if no valid correction
data in non-volatile memory and the code is 1.
Command:
PHOSPHORESCENCE DARK CURRENT SELECT
Syntax:
$DC code
Description:
Selects whether a dark current of zero or one generated by the
SET DARK CURRENT command is used. Code is 0 for dark
current = 0, 1 otherwise. Power-on default is 0.
Dark current value is subtracted from the emission main signal
on-the-fly before block averaging and ratioing (see $OS
command).
Returns current dark current value if no parameter is specified.
Содержание LS 55
Страница 1: ...LS 55 User s Guide ...
Страница 7: ...Warnings and Safety Information ...
Страница 12: ...12 LS 55 User s Guide ...
Страница 13: ...Introduction ...
Страница 19: ...Introduction 19 Supply Voltages 100 V 120 V 220 V 230 V Frequency 50 60 Hz Power 150 VA ...
Страница 20: ...20 LS 55 User s Guide ...
Страница 21: ...Unpacking and Installation ...
Страница 36: ...36 LS 55 User s Guide ...
Страница 37: ...Accessories Installation and Use ...
Страница 72: ...72 LS 55 User s Guide Figure 54 Using the powder accessory ...
Страница 93: ...Accessories Installation and Use 93 Figure 86 Masking the outer areas ...
Страница 121: ...RS232 Commands ...
Страница 163: ...System Description and Maintenance ...
Страница 165: ...System Description and Maintenance 165 Figure 112 The optical layout ...
Страница 198: ...198 LS 55 User s Guide ...
Страница 199: ...Index ...