92 .
LS 55 User’s Guide
17.
Rotate the right fibre until the excitation light beam image is centralized on
the optical fibres.
18.
Loosen the grub screw again and move the fibre into and out of the holder
(keeping the fibre bundle vertical) to optimize the focus.
19.
Tighten the knurled screw under the base plate to lock the fibre.
20.
Replace the LS 55 central optical cover.
21.
Turn on the photomultiplier voltage.
22.
Close the main cover.
Operation
The ~22
o
angle of emergence of light from the sampling end means that the
further the sampling end is from the sample, the larger the excitation image will be,
as shown in Figure 85.
Figure 85 Sampling positions
However, if the sampling end contacts the sample, very little light will be accepted
by the emission fibre. Optimal sensitivity is achieved with a 1-2 mm distance
between the fibre optic and the sample.
Improving resolution further can be achieved by masking the sampling end of the
fibre to eliminate the outer parts of the semicircles from measurement:
Содержание LS 55
Страница 1: ...LS 55 User s Guide ...
Страница 7: ...Warnings and Safety Information ...
Страница 12: ...12 LS 55 User s Guide ...
Страница 13: ...Introduction ...
Страница 19: ...Introduction 19 Supply Voltages 100 V 120 V 220 V 230 V Frequency 50 60 Hz Power 150 VA ...
Страница 20: ...20 LS 55 User s Guide ...
Страница 21: ...Unpacking and Installation ...
Страница 36: ...36 LS 55 User s Guide ...
Страница 37: ...Accessories Installation and Use ...
Страница 72: ...72 LS 55 User s Guide Figure 54 Using the powder accessory ...
Страница 93: ...Accessories Installation and Use 93 Figure 86 Masking the outer areas ...
Страница 121: ...RS232 Commands ...
Страница 163: ...System Description and Maintenance ...
Страница 165: ...System Description and Maintenance 165 Figure 112 The optical layout ...
Страница 198: ...198 LS 55 User s Guide ...
Страница 199: ...Index ...