128 .
LS 55 User’s Guide
Source Control and Photomultiplier Control
Thirty minutes after the last user command has been completed, the source is
turned off but the idle time drive continues.
The PMT voltage is set nominally to a voltage dependent on the excitation slit
position and the position of the excitation monochromator.
A PMT overload detection mechanism operates to indicate conditions where the
ordinate data may be distorted by saturation effects. An error is indicated if the
protection mechanism becomes active during a period of data collection but data
collection continues.
Slit and Filter Motors
All motors listed here may only be driven when a
scan or time drive is not active.
Excitation and emission slits are provided as standard and each may be
independently driven in the range 0 and 2.5 to 15 nm (excitation), 0 and 2.5 to 20
nm (emission) to positions that are multiples of 0.1 nm.
NOTE:
Slit settings below 2.5 nm are not calibrated.
An emission mirror is provided as an option. When the mirror is moved into place
the emission monochromator is driven to 800 nm and emission wavelength
subsequently assumed to be 0.
An emission filter with 8 positions is provided. The positions are: cut-off filter 1 to
5; blank; 1% attenuator and no filter.
An excitation cutoff filter at 390 nm is provided. This is switched in to the beam at
410 nm to remove second order artifacts.
An emission polarizer is provided as an accessory with the following positions:
horizontal polarizing filter; vertical polarizing filter; and no filter.
An excitation polarizer is provided as an accessory with the following positions:
horizontal polarizing filter, vertical polarizing filter, blank, and no filter.
A conceptual excitation shutter is provided in the form of an independent source
ON/OFF switch.
Analogue Output
Analogue outputs of 0 - 10 V and 0 - 1 V are provided with a resolution of one part
in 1024. The data output is processed in the same way as that output via the
RS232C channel. The data is multiplied by a user specified scaling factor and then a
user specified offset is added before output.
Содержание LS 55
Страница 1: ...LS 55 User s Guide ...
Страница 7: ...Warnings and Safety Information ...
Страница 12: ...12 LS 55 User s Guide ...
Страница 13: ...Introduction ...
Страница 19: ...Introduction 19 Supply Voltages 100 V 120 V 220 V 230 V Frequency 50 60 Hz Power 150 VA ...
Страница 20: ...20 LS 55 User s Guide ...
Страница 21: ...Unpacking and Installation ...
Страница 36: ...36 LS 55 User s Guide ...
Страница 37: ...Accessories Installation and Use ...
Страница 72: ...72 LS 55 User s Guide Figure 54 Using the powder accessory ...
Страница 93: ...Accessories Installation and Use 93 Figure 86 Masking the outer areas ...
Страница 121: ...RS232 Commands ...
Страница 163: ...System Description and Maintenance ...
Страница 165: ...System Description and Maintenance 165 Figure 112 The optical layout ...
Страница 198: ...198 LS 55 User s Guide ...
Страница 199: ...Index ...