LS 55 User’s Guide
Command: COLLECT
FFA
DATA
Syntax:
$XF cycles, prompt
Description:
Starts FFA data collection and output. Data points are output in
format as for $FK command, but in cycles of four, with the first
point always corresponding to the first filter position flashed after
the datum position for the currently rotating unit.
The specified whole number of cycles worth of data points are
output (range 1 - 99999). The cycle duration is as specified by
the $FI setting. Cycle lengths > 4 mains cycle, data is block
averaged. No smoothing or normalization is applied to the data
regardless of the $0$ state.
The optional prompt parameter specifies one of EVENT(0) or
START(1) input line triggers for-data collection and output.
Returns error code 0005 if no parameters given, error code 0111
if no FFA unit running, error code 0112 if synchronization error
occurs, error code 0099 if aborted
Содержание LS 55
Страница 1: ...LS 55 User s Guide ...
Страница 7: ...Warnings and Safety Information ...
Страница 12: ...12 LS 55 User s Guide ...
Страница 13: ...Introduction ...
Страница 19: ...Introduction 19 Supply Voltages 100 V 120 V 220 V 230 V Frequency 50 60 Hz Power 150 VA ...
Страница 20: ...20 LS 55 User s Guide ...
Страница 21: ...Unpacking and Installation ...
Страница 36: ...36 LS 55 User s Guide ...
Страница 37: ...Accessories Installation and Use ...
Страница 72: ...72 LS 55 User s Guide Figure 54 Using the powder accessory ...
Страница 93: ...Accessories Installation and Use 93 Figure 86 Masking the outer areas ...
Страница 121: ...RS232 Commands ...
Страница 163: ...System Description and Maintenance ...
Страница 165: ...System Description and Maintenance 165 Figure 112 The optical layout ...
Страница 198: ...198 LS 55 User s Guide ...
Страница 199: ...Index ...