System Description and Maintenance . 193
Cleaning the air filter
In normal operating conditions the air filter should be cleaned every time the
source is replaced. The air filter is situated on the right-hand side of the rear panel,
as shown in Figure 137.
Figure 137 The air filter
The air filter is cleaned as follows:
WARNING
Switch off the mains supply switch of the LS 55 and remove
the mains supply cable before starting this procedure.
1.
Switch off the LS 55 and remove the mains supply plug.
2.
Place a flat bladed screwdriver in one of the two slots on the right side of the
air filter holder and then lever off the air filter holder.
3.
Remove the air filter.
4.
Clean off dust from the filter grid.
5.
Replace the grid in the holder and refit the filter holder by snapping in to
place.
6.
Reconnect the mains cable and switch on.
Содержание LS 55
Страница 1: ...LS 55 User s Guide ...
Страница 7: ...Warnings and Safety Information ...
Страница 12: ...12 LS 55 User s Guide ...
Страница 13: ...Introduction ...
Страница 19: ...Introduction 19 Supply Voltages 100 V 120 V 220 V 230 V Frequency 50 60 Hz Power 150 VA ...
Страница 20: ...20 LS 55 User s Guide ...
Страница 21: ...Unpacking and Installation ...
Страница 36: ...36 LS 55 User s Guide ...
Страница 37: ...Accessories Installation and Use ...
Страница 72: ...72 LS 55 User s Guide Figure 54 Using the powder accessory ...
Страница 93: ...Accessories Installation and Use 93 Figure 86 Masking the outer areas ...
Страница 121: ...RS232 Commands ...
Страница 163: ...System Description and Maintenance ...
Страница 165: ...System Description and Maintenance 165 Figure 112 The optical layout ...
Страница 198: ...198 LS 55 User s Guide ...
Страница 199: ...Index ...