The advantages of this system over traditional Dewar flasks are:
•
Bubbles do not occur within the optical path since the nitrogen tank is located
below the optical plane of the instrument, leading to reduced noise.
•
Positioning of the cuvette and hence signal reproducibility are more accurate.
•
Cooling occurs more rapidly, allowing for improved sample throughput.
•
Fewer reflective and absorptive surfaces are located within the optical path.
Installation
Before installing the accessory, any filters or attenuators that need to be inserted
into the sample area brackets should be installed. Figure 108 shows the parts of
the Low Temperature Accessory.
Figure 108 Cross-section through the Low Temperature Accessory
Liquid nitrogen
Nitrogen tank
Polystyrene
insulation
Funnel
Funnel cover
Sample cover
Accessory cover
Sample tube
Foam cover
Sample chamber
Copper ‘cold finger’
Locating pin
Содержание LS 55
Страница 1: ...LS 55 User s Guide ...
Страница 7: ...Warnings and Safety Information ...
Страница 12: ...12 LS 55 User s Guide ...
Страница 13: ...Introduction ...
Страница 19: ...Introduction 19 Supply Voltages 100 V 120 V 220 V 230 V Frequency 50 60 Hz Power 150 VA ...
Страница 20: ...20 LS 55 User s Guide ...
Страница 21: ...Unpacking and Installation ...
Страница 36: ...36 LS 55 User s Guide ...
Страница 37: ...Accessories Installation and Use ...
Страница 72: ...72 LS 55 User s Guide Figure 54 Using the powder accessory ...
Страница 93: ...Accessories Installation and Use 93 Figure 86 Masking the outer areas ...
Страница 121: ...RS232 Commands ...
Страница 163: ...System Description and Maintenance ...
Страница 165: ...System Description and Maintenance 165 Figure 112 The optical layout ...
Страница 198: ...198 LS 55 User s Guide ...
Страница 199: ...Index ...