RS232 Commands . 143
Command: FILTER
SELECT
Syntax:
$FL code
Description:
Selects the filtering applied in the data processing. Code is
integer with one of following values: 0 Savitsky-Golay filter,
1 Binomial filter. Power-on default is 0.
Command: FILTER
RESPONSE
Syntax:
$FR response number
Description:
Sets the active data processing filter characteristics for
subsequent data processing. Response number must be an odd
integer in range 3 to 101. Power-on default is response 19.
Returns response of currently active filter if no parameter is
specified.
Command: PHOSPHORESCENCE
CYCLE
Syntax:
$PC time
Description:
Sets phosphorescence mode cycle time. Time is an integer in the
range 0 to 9999 and has units of ms. If a time of zero is specified
then the cycle time will be set to the minimum value required
given current data collection parameters.
Returns error 0103 if time specified cannot be accommodated
with current phosphorescence data collection parameters.
Power-on default is to set minimum cycle time.
Содержание LS 55
Страница 1: ...LS 55 User s Guide ...
Страница 7: ...Warnings and Safety Information ...
Страница 12: ...12 LS 55 User s Guide ...
Страница 13: ...Introduction ...
Страница 19: ...Introduction 19 Supply Voltages 100 V 120 V 220 V 230 V Frequency 50 60 Hz Power 150 VA ...
Страница 20: ...20 LS 55 User s Guide ...
Страница 21: ...Unpacking and Installation ...
Страница 36: ...36 LS 55 User s Guide ...
Страница 37: ...Accessories Installation and Use ...
Страница 72: ...72 LS 55 User s Guide Figure 54 Using the powder accessory ...
Страница 93: ...Accessories Installation and Use 93 Figure 86 Masking the outer areas ...
Страница 121: ...RS232 Commands ...
Страница 163: ...System Description and Maintenance ...
Страница 165: ...System Description and Maintenance 165 Figure 112 The optical layout ...
Страница 198: ...198 LS 55 User s Guide ...
Страница 199: ...Index ...