184 .
LS 55 User’s Guide
CAUTION
Excessive force will break the lamp, leaving sharp pieces in the
lamp housing.
9.
Remove the new lamp from its packaging.
CAUTION
Take care-the lamp is EXTREMELY sensitive. A slight knock
against the table surface, for example, can immediately break
the lamp. Also take care not to touch the lamp, or the central
capillary section could become contaminated leading to serious
loss of sensitivity. If the capillary is accidentally touched, wipe
immediately with spectroscopic grade ethanol or methanol and
dry with a cuvette-wipe tissue.
NOTE:
In order to seat the lamp correctly in the bracket, it needs to be rotated
before insertion into the bracket so that it is correctly aligned with the
optical system.
10.
Turn the lamp so that the trigger strip, as shown in Figure 127, is located to
the right, pointing towards the small mirror next to the entrance slit, as
shown in Figure 128.
Figure 127 Details of the electrode
Содержание LS 55
Страница 1: ...LS 55 User s Guide ...
Страница 7: ...Warnings and Safety Information ...
Страница 12: ...12 LS 55 User s Guide ...
Страница 13: ...Introduction ...
Страница 19: ...Introduction 19 Supply Voltages 100 V 120 V 220 V 230 V Frequency 50 60 Hz Power 150 VA ...
Страница 20: ...20 LS 55 User s Guide ...
Страница 21: ...Unpacking and Installation ...
Страница 36: ...36 LS 55 User s Guide ...
Страница 37: ...Accessories Installation and Use ...
Страница 72: ...72 LS 55 User s Guide Figure 54 Using the powder accessory ...
Страница 93: ...Accessories Installation and Use 93 Figure 86 Masking the outer areas ...
Страница 121: ...RS232 Commands ...
Страница 163: ...System Description and Maintenance ...
Страница 165: ...System Description and Maintenance 165 Figure 112 The optical layout ...
Страница 198: ...198 LS 55 User s Guide ...
Страница 199: ...Index ...