LS 55 User’s Guide
Changing the Instrument Operating Voltage
WARNING
Any adjustment, maintenance, or repair of the opened, operating
instrument, must be performed by a skilled person who is aware
of the hazards involved (a
RESPONSIBLE BODY
).
WARNING
Switch off the mains supply switch of the LS 55 and remove
the mains supply cable before starting this procedure.
The operating voltage of the instrument can be changed by altering the
connections on the mains transformer, located inside the instrument.
1.
Switch off the instrument and disconnect the electrical mains supply.
2.
Open the main cover as detailed in
CAUTION
Do not change any of the terminal connections on the right-
hand side of the transformer (viewed from the front of the
instrument).
The mains transformer is located at the rear of the instrument. On the left-hand
side of the transformer (viewed from the front) are five terminals with the voltages
they each correspond to marked on the top of the transformer, as shown in
Figure 138.
Содержание LS 55
Страница 1: ...LS 55 User s Guide ...
Страница 7: ...Warnings and Safety Information ...
Страница 12: ...12 LS 55 User s Guide ...
Страница 13: ...Introduction ...
Страница 19: ...Introduction 19 Supply Voltages 100 V 120 V 220 V 230 V Frequency 50 60 Hz Power 150 VA ...
Страница 20: ...20 LS 55 User s Guide ...
Страница 21: ...Unpacking and Installation ...
Страница 36: ...36 LS 55 User s Guide ...
Страница 37: ...Accessories Installation and Use ...
Страница 72: ...72 LS 55 User s Guide Figure 54 Using the powder accessory ...
Страница 93: ...Accessories Installation and Use 93 Figure 86 Masking the outer areas ...
Страница 121: ...RS232 Commands ...
Страница 163: ...System Description and Maintenance ...
Страница 165: ...System Description and Maintenance 165 Figure 112 The optical layout ...
Страница 198: ...198 LS 55 User s Guide ...
Страница 199: ...Index ...