78 .
LS 55 User’s Guide
12.
Rotate the mirror via the adjustment screw, as shown in Figure 62, until the
light image is centralized on the fibre bundle, as shown in Figure 63.
Figure 63 The end of the fibre bundle as seen in the dentist’s mirror
13.
Tighten the locking screw.
14.
Open the LS 55 main cover as detailed in
on
15.
Remove the main optical cover as detailed in
on page 176.
16.
Switch off the room lighting.
This enables you to see the emission entrance slit easier.
17.
Set the excitation wavelength to zero order.
This produces white light.
18.
Set the emission slit width to 10 nm.
19.
Place a plane mirror beneath the fibre optic sampling end.
This reflects the white light back into the instrument, for alignment purposes.
20.
Observe the emission entrance slit though the open aperture in the emission
filter wheel, as shown in Figure 64.
Содержание LS 55
Страница 1: ...LS 55 User s Guide ...
Страница 7: ...Warnings and Safety Information ...
Страница 12: ...12 LS 55 User s Guide ...
Страница 13: ...Introduction ...
Страница 19: ...Introduction 19 Supply Voltages 100 V 120 V 220 V 230 V Frequency 50 60 Hz Power 150 VA ...
Страница 20: ...20 LS 55 User s Guide ...
Страница 21: ...Unpacking and Installation ...
Страница 36: ...36 LS 55 User s Guide ...
Страница 37: ...Accessories Installation and Use ...
Страница 72: ...72 LS 55 User s Guide Figure 54 Using the powder accessory ...
Страница 93: ...Accessories Installation and Use 93 Figure 86 Masking the outer areas ...
Страница 121: ...RS232 Commands ...
Страница 163: ...System Description and Maintenance ...
Страница 165: ...System Description and Maintenance 165 Figure 112 The optical layout ...
Страница 198: ...198 LS 55 User s Guide ...
Страница 199: ...Index ...