System Description and Maintenance . 185
Figure 128 Aligning the electrode
11.
Insert the new lamp into the lamp holder bracket.
There should be little resistance, but if this is not the case, loosen off the grub
screw.
CAUTION
Use of excessive force when inserting the new lamp will most
probably break it.
12.
When the lamp has been fully inserted into the bracket, gently tighten the
grub screw.
NOTE:
The grub screw does not need to be overtight. Once the grub screw makes
contact with the lamp post, this is tight enough. Overtightening could lead
to scoring on the column of the lamp, making future removal difficult.
13.
Connect both electrode cables by depressing the connector, inserting the
cable, and releasing the connector, as shown in Figure 129.