RS232 Commands . 131
Section 2 - Command Formats
Following is a list of error codes.
0000 No
error.
CODES INDICATING END OF COMMAND EXECUTION
0002
Command not recognized.
0003
Invalid parameter format (real where integer expected).
0005
Not enough parameters.
0006
Illegal character in expected numeric parameter.
0008
Negative numeric where positive expected.
0009
Value out of range.
0014
Input buffer overflow.
0022
Parity, framing, or overrun error.
0099
Abort termination of command.
0100
Accessory not fitted.
0103
Invalid phosphorescence data collection parameters.
0104
Calibration peak not found.
0105
Calibration operation not successful - user message gives reason.
0106
Non-volatile memory error - fatal.
0107
PMT over range warning.
0108
Non-volatile memory error - recovered from backup copies.
0109
Too many parameters.
0110 Instrument
busy.
0111
Invalid parameter for instrument in current state.
0112
Motor stepping error or Fast Filter Accessory missing synchronous
pulse.
Содержание LS 55
Страница 1: ...LS 55 User s Guide ...
Страница 7: ...Warnings and Safety Information ...
Страница 12: ...12 LS 55 User s Guide ...
Страница 13: ...Introduction ...
Страница 19: ...Introduction 19 Supply Voltages 100 V 120 V 220 V 230 V Frequency 50 60 Hz Power 150 VA ...
Страница 20: ...20 LS 55 User s Guide ...
Страница 21: ...Unpacking and Installation ...
Страница 36: ...36 LS 55 User s Guide ...
Страница 37: ...Accessories Installation and Use ...
Страница 72: ...72 LS 55 User s Guide Figure 54 Using the powder accessory ...
Страница 93: ...Accessories Installation and Use 93 Figure 86 Masking the outer areas ...
Страница 121: ...RS232 Commands ...
Страница 163: ...System Description and Maintenance ...
Страница 165: ...System Description and Maintenance 165 Figure 112 The optical layout ...
Страница 198: ...198 LS 55 User s Guide ...
Страница 199: ...Index ...