Page 57
FILTER PARAMETERS
FILTER VALUE [1201]
This parameter adjusts not only the refresh rate of the display, but mostly the serial and analogue output.
The maximum refresh rate of the display is limited to 10 Hz
High filter values speed up the display update.
Low filter values slow down the display update.
Factor
(Hz)
Settling
Time
(mS)
Freq ADC
(Hz)
Reading no.
Monotony
time (mS)
Oscillation
time (mS)
Oscillation
range (div)
MANUAL [0]
Selectable
To be set by
the
To be set by
the
To be set by
the
To be set by
the
50 [1]
20
250
5
20
4000
10
25 [2]
40
100
5
40
3000
12
10 [3]
100
50
5
80
2500
16
5 [4]
200
50
10
100
2000
20
2 [5]
500
50
25
250
1500
25
1.25 [6]
800
12.5
10
300
1500
25
1 [7]
1000
12.5
12
400
1500
25
0.7 [8]
1500
12.5
19
500
1200
30
0.5 [9]
2000
12.5
25
600
1000
30
Default: 2 Hz
The following parameters are visible and therefore can be set, only if the parameter selection is MANUAL.
OUTPUT RATE [1202]
With this parameter, the frequency of weight acquisition is adjusted
Selectable values:
12.5 [0]
50 [1]
100 [2]
250 [3]
Default: it depends on the value set in the Filter Value parameter
AVERAGE NUMBER [1203]
This parameter allows you setting the number of readings that the filter will use to establish the average
value.
Value from 0 to 50.
Default: it depends on the value set in the Filter Value parameter