Page 50
COM 2 (RS485)
VALUE SENT
Selection of the value sent on RS 485 output.
Selectable values:
NET, GROSS
Default: NET
PROTOCOL
It defines how to use the RS485 serial port:
Selectable values:
NONE
Serial communication deactivated
CONTINUOUS:
Continuous sending of the weight string. It can be used, for example, to drive a
weight repeater. See details in the appropriate paragraph.
MODBUS:
MODBUS RTU protocol. See details in the appropriate paragraph.
Default: NONE
BAUD RATE
It defines the baudrate of the RS485 serial port:
The value must be set to the same value as the PC/PLC or remote display.
Selectable values:
1200, 2400, 4800, 9600, 19200, 38400, 57600, 115200
Default: 9600
FRAME FORMAT:
Frame type. In the case of a SLAVE or MODBUS protocol, it is not possible to select 7-bit data format
(E-7-2 and O-7-2):
Selectable values:
n-8-1, n-8-2, E-7-2, E-8-1, o-7-2, o-8-1
Default: n-8-1
ADDRESS:
Communication address of the serial port:
Values from 1 to 32
Default: 1