Page 46
TABLE CALIBRATION
It allows you to manually program up to 5 calibration points, in addition to the zero value. The values
correspond to those determined by the linearization procedure with sample weights. In this way, it is
possible to display the values determined automatically with this procedure, or modify and program
them according to preset values.
TABLE CALIBRATION
MESSAGE
DESCRIPTION
UNITS
Zero signal
Signal value corresponding to the zero scale.
mV/V
Value of weight
P1
Weight value corresponding to the 1st calibration
point.
Weight measurement
unit
Signal P1
Signal value corresponding to the 1st calibration point. mV/V
Value of weight
P2
Weight value corresponding to the 2nd calibration
point.
Weight measurement
unit
Signal P2
Signal value corresponding to the 2nd calibration
point.
mV/V
Value of weight
P3
Weight value corresponding to the 3rd calibration
point.
Weight measurement
unit
Signal P3
Signal value corresponding to the 3rd calibration
point.
mV/V
Value of weight
P4
Weight value corresponding to the 4th calibration
point.
Weight measurement
unit
Signal P4
Signal value corresponding to the 4th calibration point. mV/V
Value of weight
P5
Weight value corresponding to the 5th calibration
point.
Weight measurement
unit
Signal P5
Signal value corresponding to the 5th calibration point. mV/V
Zero Signal
Acquisition
Signal acquisition function, in mV/V, corresponding to
the zero scale
The values programmed to zero are not taken into account. Table calibration is automatically canceled
when a new theoretical calibration or a sample weight is performed.
Following the execution of the zero signal acquisition function, the signals in the table are recalculated.
At each signal value an offset is added, obtained from the difference between the new acquired zero
signal and the old zero signal value.