Page 20
OPERATING FUNCTIONS
SWITCHING THE GROSS/NET WEIGHT DISPLAY
Press on the weight dial on the instrument main screen to switch the weight display. Each time the di-
splayed weight changes between the following values:
• Gross - the gross weight is displayed in the weight dial.
• Net - the net weight is displayed in the weight dial.
SEMI-AUTOMATIC ZERO
This operation is performed to correct small scale zero displacements.
The command to reset the gross weight is not run under the following conditions:
• Unstable weight (the weight does not stabilize within 3 seconds of the reset command).
• Gross weight, compared to the original zero calibration, greater in positive or negative of the
value entered in the Zero Band parameter (instrument setup menu, metrological parameters). If this
parameter is set to 0, the semi-automatic zero function is disabled.
The operation to reset the gross weight is stored when the instrument is switched off.
AUTOMATIC TARE
Press on the button to run the automatic tare.
The automatic tare command is not executed under the following conditions:
• Unstable weight (the weight does not stabilize within 3 seconds of the automatic tare command).
• Gross weight equal to or greater than the maximum capacity of the weighing system.
The tare value is stored when the instrument is switched off.
DELETE TARE
This operation deletes the previously entered tare value.