Page 47
ANALOG OUTPUT PARAMETERS (OPTIONAL)
ANALOG OUTPUT RANGE [1506]
Selection of the analog output range.
Selectable value:
0÷10 VDC [0]
0÷5 VDC [1]
4÷20 mA [2]
0÷20 mA [3]
Default: 0÷10 VDC
ANALOG OUTPUT OPERATION MODE [1505]
Selection of the value to be associated to the analog output, corresponding to the net or gross weight
or to the peak value.
Selectable value:
NET [0]
GROSS [1]
Default: NET
ZERO OFFSET[1501-1502]
Analog output zero offset. If set, the zero value of the analog output corresponds to the weight value
set in this parameter.
Analog value to subtract referred to the analog output full scale.
FULL SCALE [1503-1504]
It is the weight corresponding to the full scale of the analogue output.
Value that can be set from 0 to Capacity
Default: Capacity
OUTPUT ADJUSTMENT:
This parameter adjusts the zero and full scale value of the selected analog output
so that the PLC and the display of the MCT 1302 will indicate the same weight.
- ZERO ADJUSTMENT:
Measure the analog output value with a tester to perform zero calibration (0).
Use the + and - keys to adjust the analog output. Press and hold the key for a quick change. Press the
SAVE key to store the data.
- F.S. ADJUSTMENT: [1508]
Measure the analog output value with a tester to perform full scale calibration (FS).
Use the + and - keys to adjust the analog output. Press and hold the key for a quick change. Press the
SAVE key to store the data.
N.B. The analog output is calibrated at the factory for each selectable range. This procedure is
available to the user for further adjustment, for each selectable range. In case of complete reset of
the setup memory, the factory calibration is re-established.