Checking Your Blood Glucose
81
7
3. To help you see the test strip port in reduced lighting, when
you insert a test strip into the PDM, you can turn on the port
light. Press the middle soft key labeled
Light
. To turn the light
off, press
Light
again.
4. Wait for the PDM to display “Apply a blood sample to the
strip” on the screen (Figure 7-6).
5. Gently touch only ONE EDGE of the test strip to the control
solution. Specifically, apply the control solution to the test
strip edge next to one of the dark-colored half-circles (Figure
7-7). If
BG sound
is set to
On
, the PDM beeps when the test
strip is full.
If you do not apply the sample within 10 seconds, the strip-
port light turns off. To turn it on again, press
Light
.
From some PDM screens, you cannot access the Free-
style® blood glucose meter. For example, you cannot
use the meter while you are activating a Pod or when
an alert, alarm, or communication error screen is dis-
played. In these cases, if you insert a test strip, the
PDM beeps to alert you.
The code number on the screen should match
the code number on the side of your test strip
vial (Figure 7-5 on previous page). They must
always match, or your results will be inaccurate.
(See “Set the blood glucose test strip code” later
in this chapter.)
Figure 7-6
Figure 7-7