Using the Pod
47
5
7. Turn the vial and syringe upside down. Withdraw insulin from
the vial into the syringe, expelling any air bubbles. Fill at least
to the MIN (minimum) fill line (Figure 5-9).
8. Remove the needle from the vial and insert it straight down
into the insulin fill port on the underside of the Pod
(Figure 5-10).
9. Depress the syringe plunger to completely empty the insulin
into the Pod. The Pod will beep, indicating that the System is
ready to proceed to the next step.
10. Remove the needle from the insulin fill port. The port is self-
sealing; insulin will not leak after the needle is removed.
Avoid using insulin from more than one vial,
which may introduce air into the syringe.
Figure 5-9
Figure 5-10
To ensure proper fill, do not insert fill syringe at an
angle into the fill port.
Do not use any other type of needle or filling
device besides the syringe provided with each
Pod.
NEVER use a Pod if you hear a crackling noise or
feel resistance when you depress the plunger.
These conditions can result in underdelivery of
insulin.
Do not insert the fill syringe into the fill port more
than once.
NEVER inject air into the fill port. Doing so may
result in unintended or interrupted insulin delivery.