Errors, Advisories, and Hazard Alarms
124
10
■
Advisory Alarms
Advisory alarms
tell you about a condition that requires your
attention. Each advisory alarm beeps intermittently. When you
hear an advisory alarm, first turn on the PDM to display the
Sta-
tus
screen. The PDM communicates with the Pod, if necessary,
and the screen displays a message describing the alarm and how
to resolve it.
Some advisory alarms are controlled by levels you set in the
Set-
tings
screen. One example is the level at which the
Low
reservoir volume
alarm sounds.
All advisory alarms are time-sensitive and require a response.
Some advisory alarms escalate to
hazard alarms
if you do not
respond within a certain time period (as noted in the table
below).
Alarm
Source
Beep Pattern
Action
Low reservoir.
Replace pod soon.
(This is based on the volume
level you selected.)
Pod
and
PDM
2 sets of beeps every minute for 3 minutes.
Repeats every 60 minutes until acknowledged.
Escalates to hazard alarm when empty. See
“Hazard Alarms” later in this chapter.
Press
OK
to acknowledge.
Expiration alert.
Pod expires at 5:00 pm 2/11
Pod
2 sets of beeps every minute for 3 minutes.
Repeats every 15 minutes until acknowledged.
Press
OK
to acknowledge.
Change Pod (see Chapter 5, Using the
Pod).
Pod expiration advisory.
Change pod now.
Pod
2 sets of beeps that repeat periodically.
Repeats every 15 minutes during the last hour of
the Pod’s life.
Escalates to
Pod expired
hazard alarm.
Press
OK
to acknowledge.
Change Pod (see Chapter 5, Using the
Pod).