Using the Pod
44
5
Gather equipment and supplies
Gather the following before you begin:
•
Vial of rapid-acting U-100 insulin (See the warning on page x
of the Introduction for insulins approved for use in the Omni-
Pod System.)
•
An unopened Pod
•
Alcohol prep swab
Deactivate the current Pod
1. On the
Home
screen, use the
Up/Down Controller
buttons
to choose
More actions
, then press
Select
.
2. Choose
Change Pod
(Figure 5-1), then press
Select
.
3. Press
Confirm
(Figure 5-2) to deactivate the current Pod.
NEVER use insulin that is cloudy; it may be old or inac-
tive. Failure to use rapid-acting U-100 insulin, or using
insulin that is old or inactive, may lead to hyperglyce-
mia or diabetic ketoacidosis (DKA).
If you are a first-time OmniPod System user, your
healthcare provider will guide you through the steps
for initializing and applying your first Pod. Do NOT
attempt to apply or use a Pod until you have been
trained by your healthcare provider. Use of the Sys-
tem with inadequate training or improper setup
could put your health and safety at risk.
Figure 5-1
Figure 5-2