Errors, Advisories, and Hazard Alarms
127
10
If the PDM cannot communicate with the Pod to accept or shut
off a hazard alarm, you can shut it off manually.
1. Remove the Pod from your body if you have not already
done so.
2. Press the end of a paper clip or similar item into the manual
alarm shut-off port on the top side of the Pod. (See Chapter
11, Communication Failures, Figure 11-17.)
Resources for OmniPod System users are available
from the Customer Care tab on our Web site:
MyOmniPod.com
Among these resources are an online copy of this
User Guide
and the
My OmniPod System Information
form, which you will also find at the end of this
guide.
Alarm
Source Action
Empty reservoir.
Insulin delivery stopped.
Change pod now.
Pod
and
PDM
Press
OK
to acknow-
ledge.
Change Pod.
Check blood glucose.
Occlusion detected.
Insulin delivery stopped.
Change pod now.
Pod
Press
OK
to acknow-
ledge.
Change Pod.
Check blood glucose.
Pod expired.
Insulin delivery stopped.
Change pod now.
Pod
Press
OK
to acknow-
ledge.
Change Pod.
Check blood glucose.
Auto-off.
Remove pod now.
Pod
and
PDM
Press
OK
to acknow-
ledge.
Change Pod.
Check blood glucose.