ATTENTION: Cover the baby’s eyes with protective  

  eye shields prior to initiating phototherapy. 

2.

  

Place the baby (unclothed) in the neoBLUE cozy  

  device on top of the covered mattress. 
  Note: 

The baby’s head should be positioned at the  

  curved end of the device. The baby should be  

  placed with his/her back against the mattress.

  A blanket may be placed over the baby for  

  added warmth.

Rolled up blankets, head protectors 

or bumpers can be placed within the  

device as needed. Ensure that they  

do not compromise phototherapy 

treatment to the baby. 

Ensure that 

blankets are not blocking vents.

3.

 

Turn on the phototherapy device by moving the  

  power switch to the ON “I” position.

  Note: 

If the surface temperature of the LED panel  

  becomes too warm, the phototherapy light will  

  automatically shut off as the ventilation fan continues to  

  operate. A red blinking light will alert you to check the  

  air vents and ensure that they are not obstructed. Once  

  the temperature has decreased, the blue LEDs will  

  automatically turn back on; however, the red flashing  

  light will remain on. To inactivate the blinking light,  

  power off the neoBLUE cozy system and then power  

  it back on.

  Note:

 Turn off the neoBLUE cozy light when feeding or  

  diapering the baby.

4.

 

Upon completion of administering phototherapy, turn  

  off the neoBLUE cozy device (move power switch to  
  Standby “   ” position).

1.

 

Plug the power supply into the base of the  

  neoBLUE cozy device.

 Route the power cord  

  of the neoBLUE cozy device so that it lies out- 

  side the vicinity of the baby and is plugged  

  into a grounded (3-prong) power outlet.  

  The channel in the base of the enclosure can  

  be used to route the cable to the head of  

  the device. 

2.

 

Place the neoBLUE cozy device in an  

  enclosure (e.g., bassinet). 

The bassinet’s  

  mattress may be removed prior to placing the 

  neoBLUE cozy device.

3.

 

Position the neoBLUE cozy device so that  

  the air vents are not covered by blankets or  
  clothing or positioned against obstructing  
  surfaces.

4.

 

Place the disposable mattress cover over  

  the neoBLUE cozy mattress. 

  Use only the mattress and covers supplied  

  with the device to ensure proper delivery  

  of light to the baby. 

5.

 

Place the covered neoBLUE cozy mattress  

  into the neoBLUE cozy system so that it  
  lies flat.

Operating Instructions

neoBLUE

®

 

cozy

 LED P

hotothEraPy

 o

vErviEw

The neoBLUE cozy LED Phototherapy System is a lightweight and portable device that is positioned underneath the baby to deliver phototherapy via blue light emitting diodes (LEDs). The neoBLUE cozy system consists of an  

enclosed LED light panel, a mattress, a disposable mattress cover, and a power supply. The blue LEDs emit light in the range of 400 – 550 nm (peak wavelength 450-470 nm). This range corresponds to the peak absorption of light 

by bilirubin, and is thus considered to be the most effective for the treatment of newborn jaundice.

 

   Warning! The device should not be used on its own outside of an enclosure. There is a potential for the baby to roll off the device.

    Warning! The device uses ambient air to keep the LEDs cool. Do not use the device if the ambient temperature is warmer than 30ºC (86ºF).

S

Etting

 

UP

 

thE

 neoBLUE 

cozy

 D

EvicE

neoBLUE cozy 

power cord

neoBLUE cozy 

rear view

air vents

neoBLUE cozy 

front view

mattress

disposable 

mattress 

cover

Power  

Switch

Red  

indicator 

light

a

DminiStEring

 P

hotothEraPy

 t

rEatmEnt

Note: 

If you are sensitive to the blue light, use the neoBLUE cozy device in a well-lighted area. You may also place  

bumpers, head protectors, or rolled up blankets on the neoBLUE cozy mattress, around the baby.

Head at  

curved end  

of device

Baby’s back 

against  

mattress

[Continued on Reverse Side]

Biliband

®

 Eye Protectors 

Sizes: Micro (P/N 900644)
          Premature (P/N 900643)
          Regular (P/N 900642)