U
SE
with
P
atiEnt
E
ncLoSUrES
SUch
aS
B
aSSinEtS
anD
r
aDiant
w
armErS
The neoBLUE cozy device must be placed within a patient enclosure. If blankets are used, ensure they are not blocking the vents on the exterior of the neoBLUE cozy device.
U
SE
with
an
i
ncUBator
The neoBLUE cozy device can be used in incubators under certain conditions.
Warning! Placement within an incubator: The light is not designed to be used in an oxygen enriched environment. Only use in the presence of room oxygen.
The operating temperature for the neoBLUE cozy device is 20-30
o
C. Though the device will initially operate in warmer conditions when new, use in warmer environments will shorten the
life of the LED panel. Monitor patient temperature when used in warmer environments.
S
UrfacE
a
rEa
c
ovEragE
The infant should be placed in the supine position on the neoBLUE cozy device. If additional surface area coverage is required, neoBLUE overhead lights may be used in conjunction with the neoBLUE cozy device.
neoBLUE L
ight
i
ntEnSity
o
UtPUt
c
onvErSion
c
hart
The light intensity output of the neoBLUE device is measured at Natus prior to shipment with the Ohmeda BiliBlanket
®
Meter. Because your facility may use a different photometer to measure the light intensity output, it is
necessary to understand how your reading correlates to the Ohmeda meter reading. The following table provides a light intensity output conversion chart.
P
hotometer
o
hmeda
B
ili
B
lanket
®
m
eter
J
oey
d
osimeter
m
odel
Jd-100
o
lymPic
B
ili
-m
eter
tm
i
ntensity
(
µ
W/cm
2
/nm):
35
91
aPProx
. 14 - 20
15
38
aPProx
. 4 - 8
If you have any questions, please contact Natus Medical Technical Service at 1-888-496-2887 or your authorized service representative.