The neoBLUE cozy device must be placed within a patient enclosure. If blankets are used, ensure they are not blocking the vents on the exterior of the neoBLUE cozy device.

U

SE

 

with

 

an

 i

ncUBator

The neoBLUE cozy device can be used in incubators under certain conditions. 

 

    

Warning! Placement within an incubator: The light is not designed to be used in an oxygen enriched environment. Only use in the presence of room oxygen.

The operating temperature for the neoBLUE cozy device is 20-30

o

 C. Though the device will initially operate in warmer conditions when new, use in warmer environments will shorten the 

life of the LED panel. Monitor patient temperature when used in warmer environments. 

S

UrfacE

 a

rEa

 c

ovEragE

The infant should be placed in the supine position on the neoBLUE cozy device. If additional surface area coverage is required, neoBLUE overhead lights may be used in conjunction with the neoBLUE cozy device.

 
neoBLUE L

ight

 i

ntEnSity

 o

UtPUt

 c

onvErSion

 c

hart

 

The light intensity output of the neoBLUE device is measured at Natus prior to shipment with the Ohmeda BiliBlanket

®

 Meter. Because your facility may use a different photometer to measure the light intensity output, it is  

necessary to understand how your reading correlates to the Ohmeda meter reading. The following table provides a light intensity output conversion chart.

P

hotometer

o

hmeda

 B

ili

B

lanket

® 

m

eter

J

oey

 d

osimeter

  

m

odel

 Jd-100

o

lymPic

 B

ili

-m

eter

tm

i

ntensity

 (

µ

W/cm

2

/nm):

35

91

aPProx

. 14 - 20

15

38

aPProx

. 4 - 8

If you have any questions, please contact Natus Medical Technical Service at 1-888-496-2887 or your authorized service representative. 

BiliBlanket is a registered trademark of Ohmeda Medical. Bili-Meter is a trademark of Olympic Medical. 

Biliband is a registered trademark of Nascor Pty Ltd.

1501 Industrial Road • San Carlos, CA  94070 • +1 650-802-0400, 800-255-3901 • Fax: +1 650-802-0401 • [email protected] • www.natus.com

 

 

P/N 051740A