144
■
Pod Specifications
Size:
1.6" wide x 2.4" long x 0.7" high
(4.1cm x 6.2cm x 1.7cm)
Weight (with full reservoir):
34 grams (1.2 oz)
Operating temperature range:
40° F to 104° F (5° C to 40° C)
Under normal circumstances, your body temperature will keep
the Pod well within this range.
Startup temperature:
above 50° F (10° C)
Storage temperature range:
14° F to 122° F (-10° C to 50° C)
Reservoir volume (deliverable):
200 units
Cannula insertion depth:
6.5mm
Watertight rating:
IPX8 (25 feet for 60 minutes)
Insulin concentration:
U-100
Alarm type:
Audible
Operating relative humidity range:
20–85%, noncondensing
Storage relative humidity range:
20–85%, noncondensing
Operating atmospheric pressure:
700 hPA to 1060 hPA
Storage atmospheric pressure:
700 hPA to 1060 hPA
Delivery accuracy (tested per IEC 60601-2-24):
Basal: +/- 5% at rates > 0.1 U/hr
Bolus: +/- 5% for all set values > 0.1 units
Bolus presets
7
Carbohydrate presets
36
Suspend
30 minutes to 2 hours
Low reservoir volume
indicator
10 to 50 units in 5-unit increments
Default is 10.0 units
Pod expiration alert
1 to 24 hours in 1-hour increments
Default is 4 hours
Auto-off alert
1 to 24 hours or Off in 1-hour
increments
Default is Off
PDM lock
On or Off
Default is Off
History storage
5400 records/90 days
Language