Communication Failures
134
11
3. Move across the room or to a different room. Then either:
a. Press
Discard
to begin the deactivation process.
b. Press
Retry
to repeat the attempts to communicate with
the Pod. The PDM tries to communicate with the Pod
again. If it cannot communicate, it displays the same
screen as in Figure 11-13, then finally a second “Bolus
status unknown” screen (Figure 11-14). Press
OK
to check
the Pod status or press
Discard
to begin the deactivation
process.
4. If at any point in this process the PDM can communicate
again with the Pod, you will see Figure 11-15. The message
indicates that your original instruction to cancel the bolus
was successful.
a. Press
OK
. The PDM displays the screen in Figure 11-16,
showing how much insulin was delivered before the
bolus was canceled.
b. Press
OK
again to return to the
Status
screen.
Figure 11-14
Figure 11-15
Figure 11-16