Getting Started
15
2
Enter the duration of insulin action
1. Use the
Up/Down Controller
buttons to enter the
duration
of insulin action
(how long insulin lasts in your body), then
press
Next
.
Select bolus increment and enter maximum bolus
1. Choose a bolus increment (how much each press of the
Up/Down Controller
buttons will increase or decrease a
bolus amount)—either 0.05, 0.10, 0.50, or 1.00 units of insu-
lin—then press
Next
.
2. Choose the maximum bolus you can give at one time, then
press
Next
.
Set extended bolus doses
1. To set the units for an extended bolus dose, choose
%
for per-
cent of the total bolus, or choose
Units
for units of insulin. To
turn off the extended bolus feature, choose
Off
. Then press
Next
.
Set low reservoir alert
1. Choose the level of insulin in the reservoir at which the
PDM should give you an alarm. Choose from 10 to 50 units
in 5-unit increments, then press
Next
.
The PDM now asks whether you want to activate a Pod
(Figure 2-4).
•
Press
Yes
if you are ready to activate a new Pod. See Chap-
ter 5, Using the Pod, to learn how to activate a new Pod.
•
Press
No
if you do not want to activate a Pod at this time.
The PDM displays the
Home
screen.
The duration-of-insulin-action setting is used to
calculate the amount of insulin on board (IOB) from
a previous correction bolus, when you use the sug-
gested bolus calculator.
The maximum bolus default is 10 units. Check with
your healthcare provider before adjusting these
settings.
Figure 2-4