Using the Personal Diabetes Manager
64
6
Ratios/factors/targets
When the suggested bolus calculator is turned
On
, you can view
and change all the settings using the
Ratios/factors/targets
menu option. Choose
Review all settings
, then press
Select
(Figure 6-7a ). Or choose one setting from the menu, then press
Select
(Figure 6-7b). The steps are the same whether you choose
all settings or individual ones.
1.
Target BG and correct-above value
—You can edit existing
segments and add segments, up to a total of four.
a. To edit an existing target BG time segment, choose the
segment, then press
Edit
. Use the
Up/Down Controller
buttons to choose a new start time, end time, target BG
value, and correct-above value (correction threshold).
Press
Next
after each entry.
b. To add a segment, choose
[add new]
, then press
New
.
Use the
Up/Down Controller
buttons to choose a start
time, end time, target BG value, and correct-above value
(correction threshold). Press
Next
after each entry.
c. When you have completed all the time segments you
want, press
Done
, then
Save
.
2.
Min BG for calcs
(minimum blood glucose value allowed for
calculating boluses)—Use the
Up/Down Controller
buttons
to enter a new minimum BG, then press
Enter
.
3.
Insulin to carb (IC) ratio
—You can edit existing time seg-
ments and add segments, up to a total of four.
a. To edit an existing IC ratio time segment, choose the seg-
ment, then press
Edit
. Use the
Up/Down Controller
but-
tons to choose a new start time, end time, and IC ratio.
Press
Next
after each entry.
b. To add a segment, choose
[add new]
, then press
New
.
Use the
Up/Down Controller
buttons to choose a start
time, end time, and IC ratio. Press
Next
after each entry.
When the suggested bolus calculator is turned
Off
, the
Ratios/factors/targets
option does not appear in the
Bolus/basal/calcs
menu.
Figure 6-7a