CHAPTER 3 REPAIR PROCEDURES
3.2.2 Load Cell Disassembly
17. Remove the two screws on the top of the Load Cell PCB. Carefully remove stepped
washer from rear screw hole, and lift off the PCB.
Figure 3-49. Removing screws from PCB.
Note stepped washer under rear screw.
Figure 3-50. Magnet Cover without PCB.
NOTE:
If the Load Cell PCB has to be replaced (for example, if Ramp does not give a
reading), after removing the PCB, unsolder the two Temperature Sensors, as shown in
Figure 3-51. Then remove the Temperature Sensors and set them aside. Remove the
Temperature Sensors from the new PCB. Insert and solder the Temperature Sensors
from the original PCB into the new PCB. The new PCB is now ready to re-install.
Solder points for top
Temperature Sensor
Top Temperature
Sensor
Solder points for
bottom Temperature
Sensor
Bottom Temperature
Sensor
Figure 3-51. Solder points on top
of Load Cell PCB.
Figure 3-52. Solder points on
bottom of Load Cell PCB.
Discovery Balance Service Manual
3-27
Ohaus Corporation www.ohaus.com