CHAPTER 2 DIAGNOSIS
TABLE 2-5. BALANCE CANNOT CALIBRATE INTERNALLY
SYMPTOM PROBABLE
CAUSE
REMEDY
Balance can be turned on
but will not calibrate using
internal calibration.
Balance may be
unstable.
The InCal Weight
Hardware may be dirty,
mis-aligned or
defective.
Calibration Motor
Assembly may be
defective
Main PC Board may be
defective.
Load Cell PCB is
defective
Level the balance using leveling
feet on rear of chassis, together
with leveling bubble, just inside
the left Door Side Draft Shield.
Disassemble the InCal Weight
Hardware and test it.
->InCal Weight Hardware
Disassembly & Testing 3.1.4.
If there are problems with the
Weight Arm movements, check
the Calibration Motor, and
replace if necessary.
->Removing & Replacing
Calibration Motor Assembly
3.1.5.
Verify that all other functions
are operational. If not, the Main
PC board may be defective.
->Main PCB Removal &
Replacement 3.1.3.
Replace Load Cell PCB.
(See p. 3-27.)
Discovery Balance Service Manual
2-5
Ohaus Corporation www.ohaus.com