APPENDIX C SOFTWARE SERVICE TOOL INSTRUCTIONS
C.2 To Replace a Load Cell:
1. Follow the steps in Section 3-2 to replace the defective load cell.
2. Open the packet containing labels that was supplied with the replacement Load Cell.
These labels each contain a model number followed by the IDNR number.
Carefully select the label that matches exactly the model number of the balance.
3. Put the new label on the balance.
4. Power the balance from an AC adapter.
5. Connect the balance to your computer.
6. Start the Discovery Service Tool Software.
7. Click on the tab labeled
Replace Loadcell
.
Figure C-3. The Software Service Tool’s Replace Load Cell tab.
8. Enter the IDNR number (xxx.xx.xx.xxxx) from the label on the back of the balance.
9. Click
the
Start
button.
10. The software will indicate the download progress. When complete, disconnect the
power from the back of the balance, then re-connect.
11. Perform Service Calibrations (see Appendix B). Test the balance.
Discovery Balance Service Manual
C-3
Ohaus Corporation www.ohaus.com