CHAPTER 3 REPAIR PROCEDURES
3.2.2 Load Cell Disassembly
There are two fine connecting wires called Wire Conductor Strips soldered to the Load Cell’s
PCB and to a smaller PCB on the top-front of the Ratio Beam. These wires must be
carefully
disconnected from the Load Cell’s PCB in order to further disassemble and repair the Load Cell.
12. Unsolder the wires at the solder points on either side of the set screw atop the Load
Cell’s PCB.
13. Remove the solder with a solder sucker, to clear holes in PCB so the wires can pass
through when you need to solder them back in place.
14. Use tweezers to remove wires from holes.
NOTE:
The smooth arc of the connecting wires is critical for the balance to function
properly. Carefully observe this arc, and be sure to replicate it when re-
soldering the wires!
Solder points
Figure 3-45. Solder points on either side of set screw
connecting fine wire to Load Cell PCB.
Discovery Balance Service Manual
3-25
Ohaus Corporation www.ohaus.com